Physical:
This product consists of
one PCB, together with two interface cables.
Cables: One D-sub 25 pin input cable and one four pin RJ11 output cable.
Enclosure: Plastic case, 4.8" x 2.6" x 1.7"
PCB size: 2.8" x 1.9"
Weight: 9 oz.
Communication Port:
RS232 to Vasotrac
APM205A
RJ11 Plug to Host IBP Patient Monitor
Requires a DC Excitation voltage from the host IBP patient monitor
from 4 to 8 volts DC.
IBP Port:
The NIA V-Line is
capable of operating with an excitation voltage from the IBP monitor from 4 to
8 volts DC. The pressure output is capable of indicating levels from 0 to 300
mmHg. One green LED indicates
power on. Supplied by the Vasotrac
APM 205A, +5VDC. Operating temperature:
10° to 40°C Vasotrac APM 205A System
with application code Version 7.13 or greater, Host IBP Patient Monitor, and
an extension transducer cable that is compatible to the Host IBP Patient
Monitor's Channel which should have an available four-conductor (RJ11) phone
jack connector end that would be used for an Arterial Line Set. These
extension cables are typically provided by the host IBP Patient Monitoring
Company. Normally, an invasive
arterial pressure module/monitor accepts analog signals from an Arterial Line.
The module/monitor channel furnishes the DC excitation (supply) voltage for
the Arterial Line's pressure transducer. The Arterial Line's pressure
transducer returns a voltage indicating the pressure present on the face of
the transducer's sensing area in the patient's artery. The NIA V-Line is
compatible with any invasive arterial blood pressure channel that provides a
transducer DC excitation (supply) voltage in the range of 4 to 8 volts DC. The Medwave NIA V-Line
Interface Package generates an analog signal that is interpreted by the IBP
monitor as if it were an Arterial Line pressure transducer signal in the
patient's artery. Application Code Version
7.13 or greater. The monitor must have an
IBP channel that utilizes DC excitation (supply) voltage in the range of 4 to
8 volts DC. The NIA V-Line will not function properly if the IBP channel
utilizes AC excitation. The host IBP patient
monitor must be equipped with a compatible extension transducer cable that has
an available four-pin conductor (RJ11) phone jack connector end, which would
otherwise be used for an Arterial Line set. These extension cables are
typically provided by your IBP patient monitoring company. +/- 2.5% of true reading
over a range of 20mmHg to 240 mmHg
Displays:
Electrical:
Note: There is an isolation barrier within the NIA V-Line that furnishes
isolation between the Vasotrac APM 205A and the IBP monitor. This barrier is
designed to provide double insulation at 240 volts.
Environmental:
Humidity: 15% - 90% non-condensing
Storage temperature: -20° to 50°C
Required Accessories:
Theory of Operation:
Equipment Requirements:
Vasotrac APM 205A Requirements:
IBP
Monitor Requirements:
Performance Specifications:
| NIA V-Line Technical Specifications |