Legato
Module Specifications

  Download / View a Printable Legato OEM Module Brochure



PHYSICAL:
This product consists of a processing PCB along with a patient cable with pressure sensor. The module interfaces to a host monitor via a serial data port.

Enclosure: None. This product is an OEM module intended for integration into the Developer’s monitoring system

PCB size: 85mm (3.35") (W) x 110mm (4.33") (L), not to exceed 35mm (1.38") in height

Weight: PCB Assembly = 2.5 oz., Patient Cable with sensor = 9.7 oz


COMMUNICATION PORT:
Three wire serial RS232: TX DATA, RX DATA, GND

RS232 voltage levels
Vout swing = +/- 5 v minimum, +/- 8 v maximum
               with outputs loaded 3 kohms to ground.
Vin high threshold = 1.8 v typical, 2.4 v maximum
Vin low threshold = 1.3 v typical, .8 v minimum

Optional TTL levels
Vout high = 4.5 v minimum @ Ioh = -200 uA,
                     3.5 v minimum @ Ioh = -1mA
Vout low = 0.4 v maximum @ Iol = 1.6mA,
                     1.0 v maximum @ Iol = 10mA
Vin high = 2.0 v minimum, 5.3 v maximum
Vin low = 0.5 v maximum , -0.3v minimum
Speed
9600 baud, 8 byte, non parity, 1 stop byte.

DISPLAYS:
None. All display tasks are handled by the host.


ELECTRICAL:
DC supply requirements: 300mA @ +5v, 150mA @ -5v


ENVIRONMENTAL:
Operating temperature: Printed Circuit Board Assembly, 10o to 70o C
                                  Sensor and patient cable, 10o to 40o C
Humidity: 15% - 90%, non-condensing
Storage temperature: -20o to 50o C


ALARMS:
System diagnostic alarms are communicated to the host via the serial port.
Patient alarms are the responsibility of the host.


THIS IS AN INVESTIGATIONAL DEVICE.
Limited by United States law to investigational use only.
IT IS NOT INTENDED FOR PATIENT DIAGNOSTIC USE.